[sheet: Device Comparison]

Parameter

--- DEVICE IDENTIFICATION ---

Device Trade NameSolstice Lumina™ CGM SystemNovaSense GlucoTrack NX-500 CGM SystemNovaSense GlucoTrack NX-700 CGM SystemSolstice GlucoStrip RX-200 Test Strip Reader
Device ComponentsLumina C14 (subcutaneously implantable CGM sensor) + Lumina R7 (wearable reader/transmitter)Implantable CGM sensor + Handheld receiverImplantable CGM sensor + Handheld receiverTest strip reader (standalone)Lumina system comprises two components submitted as a single 510(k). GlucoStrip RX-200 is a standalone device cited for corporate pedigree only — not a predicate for substantial equivalence.
ManufacturerSolstice Biomedical Inc., 220 Westbrook Drive, Suite 400, Cambridge, MA 02142 (U.S. subsidiary of Solstice Biomedical Devices Ltd., Ireland, CRO No. 648231)NovaSense Medical Technologies Inc., 4800 Innovation Parkway, San Diego, CA 92121 (Delaware File No. 5429876)NovaSense Medical Technologies Inc., 4800 Innovation Parkway, San Diego, CA 92121Solstice Biomedical Inc., 220 Westbrook Drive, Suite 400, Cambridge, MA 02142
510(k) NumberPending (target submission April 30, 2026)K211034K230489K192847
510(k) Clearance DateTarget: Q3 2026June 18, 2022March 7, 2024November 14, 2019

--- CLASSIFICATION ---

Product CodeQKQ (anticipated)QKQQKQNBWGlucoStrip RX-200 classified under product code NBW; not a predicate for substantial equivalence for the Lumina CGM System.
Device ClassClass II (anticipated)Class IIClass IIClass IIAll devices are Class II. Subject device classification to be confirmed via Pre-Sub meeting with CDRH.
Regulation Number21 C.F.R. § 862.1355 (anticipated)21 C.F.R. § 862.135521 C.F.R. § 862.135521 C.F.R. § 862.1345GlucoStrip RX-200 regulated under a different CFR section; not a predicate.
Submission TypeTraditional 510(k) (single submission covering both C14 and R7 as a system)Traditional 510(k)Traditional 510(k)Traditional 510(k)

--- INTENDED USE / INDICATIONS FOR USE ---

Intended UseContinuous glucose monitoring in persons aged 18 and older with Type 1 or Type 2 diabetes mellitusContinuous glucose monitoring in persons with diabetes mellitusContinuous glucose monitoring in persons with diabetes mellitusIn vitro quantitative measurement of glucose in capillary whole blood (test strip-based)Subject device intended use is substantively equivalent to both predicates. GlucoStrip RX-200 has a fundamentally different intended use.
Target Patient PopulationAdults aged ≥18 years with Type 1 or Type 2 diabetes mellitusAdults with diabetes mellitusAdults with diabetes mellitusAdults with diabetes mellitusSubstantially equivalent intended patient population across subject device and both predicates.
Prescription / OTCPrescription use (Rx)Prescription use (Rx)Prescription use (Rx)OTC

--- SENSOR TECHNOLOGICAL CHARACTERISTICS ---

Sensor TypeSubcutaneously implantable, enzymatic electrochemical biosensor (fourth-generation)Subcutaneously implantable, enzymatic electrochemical biosensorSubcutaneously implantable, enzymatic electrochemical biosensorN/A (test strip-based)Same fundamental sensor technology (enzymatic electrochemical) across subject device and both predicates.
Sensor Wear Duration14 days (336 hours)10 days14 daysN/ASubject device sensor wear duration (14 days) differs from primary predicate (10 days) but matches secondary predicate NX-700. Extended wear is supported by performance and biocompatibility data.
Electrode MaterialPlatinum-iridium (proprietary PtIr-4G coating)PlatinumPlatinumN/AKey difference from primary predicate. PtIr-4G uses a different coating architecture. Differences do not raise new questions of safety or effectiveness; performance testing and biocompatibility data demonstrate equivalence.
Measurement Range40–400 mg/dL40–400 mg/dL40–400 mg/dL20–600 mg/dLSubject device measurement range identical to both predicates.
Warm-Up Time≤ 60 minutes≤ 60 minutes≤ 60 minutesN/AEquivalent warm-up time.
CalibrationFactory-calibrated (no user calibration required)Factory-calibratedFactory-calibratedN/AEquivalent calibration approach.

--- WIRELESS COMMUNICATION ---

Wireless ProtocolBluetooth Low Energy (BLE) 5.2Bluetooth Low Energy (BLE) 5.0Bluetooth Low Energy (BLE) 5.0N/ASubject device uses BLE 5.2 vs. BLE 5.0 in both predicates. BLE 5.2 is backward-compatible; difference is incremental and does not raise new questions of safety or effectiveness.
Data Transmission RangeUp to 10 meters (line of sight)Up to 10 meters (line of sight)Up to 10 meters (line of sight)N/AEquivalent.

--- READER / DISPLAY ---

Reader Form FactorArm-worn integrated reader/transmitter (Lumina R7)Handheld receiverHandheld receiverHandheld test strip readerKey difference from both predicates. Lumina R7 is arm-worn rather than handheld. Different form factor does not change the fundamental function (receiving, processing, and displaying glucose data). User studies and human factors testing address usability.
Display FeaturesReal-time glucose values, trend arrows, 8-hour trend graphs, configurable hypo/hyperglycemia alertsReal-time glucose values, trend arrows, trend graphsReal-time glucose values, trend arrows, trend graphs, configurable alertsGlucose value (numeric display)Subject device display features are equivalent to or enhanced versions of predicate features.
Alert Latency≤ 5 minutes from sensor reading to user notification≤ 5 minutes≤ 5 minutesN/AEquivalent alert latency.
Mobile App ConnectivitySolstice Lumina™ Mobile App (iOS and Android) via BLECompanion mobile app via BLECompanion mobile app via BLEN/AEquivalent mobile connectivity approach.

--- SOFTWARE ---

Sensor FirmwareLumina Firmware v3.4.2ProprietaryProprietaryN/ADifferent firmware; functionally equivalent purpose.
Reader Operating SystemLumina Reader OS v2.1.0ProprietaryProprietaryN/A
Glucose Prediction AlgorithmLuminaPredict v2.1 (Solstice proprietary)NovaSense proprietary algorithmNovaSense proprietary algorithm (updated)N/AKey difference. Subject device uses a different proprietary glucose prediction algorithm. Algorithm performance validated through clinical study (MARD and error grid analysis).
Mobile App VersionLumina Mobile App v1.0.3ProprietaryProprietaryN/A
Software Level of ConcernMajorMajorMajorModerateSubject device software level of concern consistent with both CGM predicates.

--- PERFORMANCE SPECIFICATIONS ---

MARD Target≤ 9.0% (overall)≤ 9.5% (overall, per cleared labeling)≤ 9.0% (overall, per cleared labeling)N/ASubject device MARD target of ≤ 9.0% is consistent with secondary predicate and meets current FDA expectations for iCGM classification per FDA guidance (September 2023).
Consensus Error Grid (Clarke)≥ 99% of paired data points in Zones A+B≥ 98.5% in Zones A+B≥ 99% in Zones A+BN/ASubject device target consistent with or exceeds predicate device performance.
Sensor Functional Life336 hours (14 days × 24 hrs/day)240 hours (10 days × 24 hrs/day)336 hours (14 days × 24 hrs/day)N/ASubject device functional life matches secondary predicate. Exceeds primary predicate by 4 days; supported by performance and biocompatibility data.

--- BIOCOMPATIBILITY ---

Biocompatibility StandardsISO 10993-1:2018 and applicable sub-partsISO 10993-1 and applicable sub-partsISO 10993-1 and applicable sub-partsISO 10993-1 (limited scope)
C14 Sensor Biocompatibility TestsCytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), Irritation (ISO 10993-23), Acute Systemic Toxicity (ISO 10993-11), Subchronic Toxicity (ISO 10993-11), Implantation (ISO 10993-6), Hemocompatibility (ISO 10993-4), Genotoxicity (ISO 10993-3)Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Implantation, Hemocompatibility, GenotoxicityCytotoxicity, Sensitization, Irritation, Systemic Toxicity, Implantation, Hemocompatibility, GenotoxicityN/AComprehensive biocompatibility testing scope for C14 sensor consistent with predicate testing scope and FDA biocompatibility guidance. Testing performed by Whitmore Laboratories Inc. (A2LA Certificate No. 3847.01).
R7 Reader Biocompatibility TestsCytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), Irritation (ISO 10993-23)Cytotoxicity, Sensitization, IrritationCytotoxicity, Sensitization, IrritationCytotoxicity, Sensitization, IrritationSkin-contacting component testing consistent with predicate approach.
Biocompatibility Testing LabWhitmore Laboratories Inc., 1625 Shady Grove Road, Suite 300, Gaithersburg, MD 20878Not publicly disclosedNot publicly disclosedNot publicly disclosed

--- EMC / ELECTRICAL SAFETY ---

EMC StandardIEC 60601-1-2 (4th edition)IEC 60601-1-2IEC 60601-1-2IEC 60601-1-2Equivalent EMC testing standard.
Electrical Safety StandardIEC 60601-1 (3rd edition + amendments)IEC 60601-1IEC 60601-1IEC 60601-1

--- CYBERSECURITY ---

Cybersecurity Framework

--- REGULATORY CERTIFICATIONS ---

EU MDR CE MarkingYes — EU MDR Certificate No. MCA-2024-07319, issued September 12, 2024 by Meridian Conformity Assessment GmbH (NB 0483)CE marked (MDD/MDR transition)CE marked (MDR)CE marked (MDD)Subject device holds current EU MDR CE certification.
ISO 13485:2016 CertificationYes — certified by Meridian Conformity Assessment GmbH (NB 0483)YesYesYes
EU Commercial AvailabilitySince September 2024Available in EUAvailable in EUAvailable in EU

--- CLINICAL DATA ---

Pivotal Clinical StudyProtocol SBD-LUM-CL-001 ('LUMINA-US'); prospective, single-arm, multicenter; 350 subjects (175 T1D + 175 T2D); 6 U.S. clinical sitesProspective multicenter accuracy study (per 510(k) summary)Prospective multicenter accuracy study (per 510(k) summary)N/AClinical study designed to demonstrate equivalent or superior performance to predicate devices. CRO: Veltrano Clinical Research Organization Inc. IRB: Clearwater Institutional Review Board LLC (Reg. No. IRB-2019-04821).
Clinical Study Primary EndpointMARD ≤ 9.0%; ≥ 99% Zones A+B on Clarke Error GridMARD; Consensus Error GridMARD; Consensus Error GridN/A
IDE StatusAnticipated NSR exemption under 21 C.F.R. § 812.2(c)(3); to be confirmed via Pre-SubIDE exempt (NSR)IDE exempt (NSR)N/A

--- CORPORATE PEDIGREE ---

Prior FDA Clearance History

[sheet: Key Differences Summary]

Difference No.ParameterSubject Device — Solstice Lumina™ CGMPrimary Predicate — NovaSense GlucoTrack NX-500 (K211034)Secondary Predicate — NovaSense GlucoTrack NX-700 (K230489)Substantial Equivalence Rationale
1Sensor Wear Duration14 days (336 hours)10 days (240 hours)14 days (336 hours)Subject device wear duration aligns with secondary predicate (K230489). Extended wear validated through biocompatibility testing (ISO 10993 — including subchronic toxicity and implantation per ISO 10993-6 and -11) and clinical accuracy data over full 14-day wear period. No new safety or effectiveness questions raised.
2Electrode MaterialPlatinum-iridium (PtIr-4G proprietary coating)PlatinumPlatinumDifferent electrode coating architecture. PtIr-4G technology uses a platinum-iridium alloy with a proprietary fourth-generation enzyme layer deposition process. Biocompatibility of PtIr-4G demonstrated through full ISO 10993 testing program at Whitmore Laboratories Inc. Performance equivalence demonstrated through bench testing (linearity, precision, interference) and clinical MARD data. The difference in electrode material does not raise new questions of safety or effectiveness.
3BLE VersionBluetooth Low Energy 5.2Bluetooth Low Energy 5.0Bluetooth Low Energy 5.0BLE 5.2 is an incremental, backward-compatible update to BLE 5.0. Provides improved throughput and coexistence but does not alter fundamental communication architecture. EMC testing per IEC 60601-1-2 confirms compliance. No new safety or effectiveness questions raised.
4Reader Form FactorArm-worn integrated reader/transmitter (Lumina R7)Handheld receiverHandheld receiverDifferent physical form factor for the reader component. Lumina R7 is worn on the upper arm rather than carried as a handheld device. Fundamental function (receiving, processing, and displaying CGM data) is equivalent. Human factors/usability testing and biocompatibility testing (skin contact) address the form factor change. No new safety or effectiveness questions raised.
5Glucose Prediction AlgorithmLuminaPredict v2.1 (Solstice proprietary)NovaSense proprietary algorithmNovaSense proprietary algorithm (updated)Different proprietary algorithm. LuminaPredict v2.1 uses a different signal processing and glucose prediction methodology. Algorithm performance validated through clinical study (Protocol SBD-LUM-CL-001) with MARD target ≤ 9.0% and ≥ 99% Zones A+B on Clarke Error Grid. Software documentation per IEC 62304 and FDA software guidance. No new safety or effectiveness questions raised.

[sheet: Device Registry]

Device NameRole in Analysis510(k) NumberClearance DateProduct CodeRegulation NumberDevice ClassManufacturerManufacturer Address
Solstice Lumina™ CGM System (C14 + R7)Subject DevicePendingTarget Q3 2026QKQ (anticipated)21 C.F.R. § 862.1355 (anticipated)Class II (anticipated)Solstice Biomedical Inc. (Delaware File No. 7834291); U.S. subsidiary of Solstice Biomedical Devices Ltd. (Ireland, CRO No. 648231)220 Westbrook Drive, Suite 400, Cambridge, MA 02142
NovaSense GlucoTrack NX-500 CGM SystemPrimary PredicateK211034June 18, 2022QKQ21 C.F.R. § 862.1355Class IINovaSense Medical Technologies Inc. (Delaware File No. 5429876)4800 Innovation Parkway, San Diego, CA 92121
NovaSense GlucoTrack NX-700 CGM SystemSecondary Predicate (Reference Device)K230489March 7, 2024QKQ21 C.F.R. § 862.1355Class IINovaSense Medical Technologies Inc.4800 Innovation Parkway, San Diego, CA 92121
Solstice GlucoStrip RX-200 Test Strip ReaderReference Device (Corporate Pedigree Only — Not a Predicate)K192847November 14, 2019NBW21 C.F.R. § 862.1345Class IISolstice Biomedical Inc.220 Westbrook Drive, Suite 400, Cambridge, MA 02142